Area velocity sensor
The submersible area velocity sensor uses advanced Pulsed-Doppler technology to directly evaluate velocity profiles. The sensor measures both flow velocity and water level in order to calculate total flow volume.
With its dimensions of 15 x 25 x 108 mm, it is the smallest sensor of this type currently available. Its low profile means less interference, which results in more accurate velocity measurements, especially in low-flow situations.
Technical specifications
- Frequency : 1 MHz
- Range:
Velocity ± 5.3 m/s
Water level 0.04-1.3 m
expandable with external 4-20 mA level sensor - Accuracy:
‒ Velocity
± 0.03 m/s from -1.5 m/s to +1.5 m/s
±1 % of reading from -5.2 bis -1.5 m/s and +1.5 bis +5.2 m/s
‒ Level
± 0.5% FS (1.5 m)
‒ Flow
typically ± 2 %, depending on site conditions - Cable length: 10 m incl. (max. 80 m)
- Particle concentration: > 50 ppm
- Material: Epoxy
- Protection class : IP68 (48 h at 50 kPa, NEMA 6P)
- Approval: ATEX Zone 1; II 2G EEx e ib IIB T (option for stationary Q-Eye PSC)
- Dimensions: 108 x 25 x 15 mm (L x W x H)
Related products

Q-Eye PSC Portable
Système Doppler à impulsion mobile
Données techniques
- Capteur :
1 x vitesse d’écoulement
1 x niveau d’eau - Fréquence : 1 MHz
- Nombre de cellules : jusqu’à 18 cellules
- Plage de mesure :
vitesse d’écoulement ± 5,3 m/s
niveau d’eau 0,04-1,3 m
extensible par capteur externe 4-20 mA - Écart de mesure :
‒ vitesse d’écoulement
± 0,03 m/s de -1,5 m/s à +1,5 m/s
± 1% de la lecture de -5,2 à -1,5 m/s et de +1,5 à +5,2 m/s
‒ niveau d’eau
± 0,5% FS (1,5 m)
‒ débit
typiquement ± 2%, en fonction de la conditions locales - Affichage : 4 lignes, 20 signes
- Clavier : 4 touches
- Mémoire : carte micro SD 16 GB
- Interfaces : Wireless LAN, 4G (LTE) / 3G (HSPA+) / 2G
- Entrées : max. 2 x 4-20 mA
- Alimentation : 2 x batteries rechargeables, échangeable pendant le fonctionnement,
24 V DC +/- 10% (option) - Boîtier de protection : IP67
- Boîtier : résine synthétique HPX®